Product Detail
Oil Voltage Tester,Easy operation, high autoimmunization,high quality,and best price Introduction: IIJ-II Oil Breakdown Voltage Tester is produced by our company, which adopts industrial single plate integrated circuit, new-style I/O jack, with unique inspection and anti-jamming technology to greatly improve its reliability and functionality. Main features: Easy operation, high autoimmunization (only press button once and all work done automatically) Menu management, data input simple and convenient It can store 99 data of oil sampling tests, the test data can be reviewed and printed in 100 years even after power is off. The clock in going on after power off. The software can be upgrade through RS232 terminal. Strong anti-jamming, high precision testing. Small measurement, light weight, and convenient for outdoor operation.

Technical parameter

Power supply: AC220V ±10%, 50Hz Output voltage: 80kV, 100kV Capacity: 1.6kVA, 2.0kVA Speed of voltage increase: about 2kV/S Voltage inspection accuracy: 3% Breakdown sensitivity: <2kV Waveform distortion: ≤3% Breakdown response time: ≤10ms Working environment: temperature: 0°C~40°C, humidity: max relative humidity 85% Storage environment: temperature: -20°C ~ 60°C, humidity: max relative humidity 75% Work height: <1500m (if over 1500m, it can be customized)

Caution:

(1)  Please read the operation manual carefully before operating

(2)  Please check if the tester is well connected to earth before testing

(3)  It is forbidden to move or lift the testing cover in the coarse of testing to avoid injury by high voltage. Power must be off before replace sampling oil.

(4)  Handle with care when taking off or closing high voltage testing cover!

(5)  If the tester works abnormally after insulating oil breaks down, please power off the tester for 10 seconds, then restart again.

(6)  After printing paper is exhausted, please refer to printer explanation part (or manual appendix) to replace printing paper to avoid damage of printer head.

(7)  Keep the tester far from moisture, dust and other corrosive materials, and keep it far away from high temperature sources.

(8)  Handle with care in transportation, don’t put it up side down.

Remarks: The manual may revise accordingly without further notice in advance. If any query, please contact us.

  (For reference only)

1.  The standard oil sampling cup and the stirring nut must be kept clean and dry.

2.  The oil sampling cup and stirring nut should be dried before use. When drying them, the temperature should not be high, or else it’s easy to damage sealing gasket of oil cup and plastic stirring nut.

3.  Please put the stirring nut into oil sampling cup, fill in oil to 1/3 cup and then slightly stir the oil in the cup so as to clean the nut and cup. It is suggested to clean them twice or three times before testing.

4.  Fill sample oil into oil cup, to the height of 8-12mm to the top.

5.  Clean oil outside of cup with clean and soft cloth or paper.

6.  Put the oil sampling cup into high voltage chamber carefully. Keep the hands of the operator clean or wear clean gloves.

7.  After oil is put into oil cup, you’d better test it before long. The oil to be tested exposed in open air for long time may affect accuracy of test results.

 Install printer’s paper/replace color ribbon

1.  Printing paper installation

To start the device to display logo, press (Menu) button to enter main menu. Main menu has three options: “Set Time” “Set Parameter ” and “Inquire Data ”. To press [­ ] and [¯] button to move shadow cursor to “Inquire Data ”, press (Enter) button to enter data management status, LCD will display “Inquire No.: XX”. Please cut the head end of new paper roll in a shape of triangle. Put down front cover of printer, hold the spring clips which fixes the head of printer, pull out head about 2cm, press the triangle button of the printer for 1 second, the printing paper begins to run, push the front end of paper into inlet, the paper will be rolled into the printer. When paper goes through the head of printer, switch off the power. Hold the two sides of paper axis slightly, install printing paper on the paper axis, get the head of paper go through paper outlet, close front cover.

2.  Ribbon replacement

After the ribbon being used for a period of time, the printed letters will be weak or unclear, you have to replace ribbon at the moment.Open front cover of printer, hold the spring clips at the two sides of pulling board, pull the board outwards till ribbon box comes out, take off used ribbon, and replace a new one.

Notice: Get the printing paper go through between ribbon box and ribbon.