Size and Weight | | Name of the monitor | Multi-parameter Monitor | Size | 258mm×210mm×175mm | Weight | 4.5Kg |
| System output | | Display type | 12.1″LCD color display screen, with a resolution of 800×600 | Interface with central unit | Network port |
| Power supply | | Power Voltage | AC 100V~240V 50/60Hz | Power Input | ≤70VA | Fuse | 1.6A | Safety Level | Class I , Type CF |
| Battery | | Type | Rechargeable Lead acid cell, 12V/3.2AH | Charge time | ≤6 hours | Working hours when charged fully | ≥30min when using normally |
| Environment | | Ambient Temperature | Working 0~+40ºCTransportation and storage –20~+50ºC | Relative humidity | Working ≤85%Transportation and storage ≤93% | Atmospheric pressure | Working 860~1060 hPaTransportation and storage 500~1060 hPa |
| ECG | | Lead Mode | 5-leads ECG input3-leads ECG input | Lead selection | I, II, III, aVR, aVL, aVF, V-I, II, IIII, II, III, aVR, aVL, aVF, V1~V6 | Gain | 2.5mm/mV, 5mm/mV, 10mm/mV, 20mm/mV, 40mm/mV | Input impedance | ≥5.0 Mohm(d.c. ~ 60 Hz) | CMRR | DIAG ≥85dBMON ≥105dBOPS ≥105dB | Frequency response | DIAG 0.05~75HzMON 0.5~25HzOPS 1~15Hz | Electrode offset potential | ±300 mV d.c. → ±600mV d.c. | Leakage Current | <10 uA | ECG signal range | ±6.0 mV | Baseline recovery | <5 s After Defi. (MON or OPS mode) | Insulation | Breakdown Voltage 4000VAC 50Hz/60Hz | Indication of electrode separation | Available | Sweep speed | 12.5mm/s, 25mm/s, 50mm/s |
| HR Measuring and Alarm | | Range | 10~300 bpm | Refreshing time | Per pulse | Resolution | 1 bpm | Accuracy | ±1% or ±1 bpm, whichever is greater | Alarm range | 10~300 bpm, continuously adjustable between high limit and low limit | Alarm indication | Sound and light alarming |
|
|