Anesthesia Machine
Anesthesia Machine's
based on ergonomics of the design, structure concise and fluent, mobile, lightweight comfortable. Microcomputer control of the ventilator imported sensors and components, and production processes to ensure that rigorous internal quality products, folding bellows breathing mechanical ventilation system so that security situation, intuitive. High-precision anesthesia evaporators to ensure that the concentration of anesthesia output accuracy. Power transmission and gas metering system to ensure the safe and reliable gas supply. The aircraft brought together a number of modern technologies, is a fine performance, functionally complete anesthesia machine.
This machine uses in respiratory control , respiratory intensive care , 8.4-inch high clear display, pressure – time wave and Velocity-time wave to monitor respiratory condition Multi-directionally;
The high-precision evaporator has the stable output concentration with pressure, temperature, and flow automatic compensation; Choose one among Enflurane, isoflurane and sevoflurane for a matching;
Main technical parameters
Part of Anesthesia Machine
Console
high-strength engineering plastic rack, light, beautiful and
corrosion resistance
Scope adult, child
Gas source O2,N2O,Air:0.3~.5MPa
flow meter O2: 0.1~10L/min 1.1~10L/min
N2O: 0.1~10L/min 1.1~10L/min
Air: 0.1~10L/min
O2,N2O linkage and nitrous oxide Truncation device
When use nitrous oxide, oxygen concentration>25%; when the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
Rapid oxygen flow 35~75L/min
Vaporizer it has the function of pressure, temperature, and flow automatic compensation. The regulation range of evaporator concentration is 0~5vol%. Choose two among enflurane, isoflurane and sevoflurane for a matching according to the demands of customer. We also prepare original imported Penlon and Drager vaporizers for choosing.
Low oxygen pressure alarm
it will be alarmed when the oxygen pressure < 0.2MPa
Respiratory circuit
work mode all close. semi-close. semi-open
APL 0.5~7KPa
Bellows breathing
adults and child bellows tidal volume range:0~1500ml
child bellows tidal volume range:0~300ml
adults,children bellows,children loop are arbitrary choice for users .
Part of Ventilator
Display mode 8.4-inch high clear display
Ventilation mode IPPV SIPPV SIMV PEEP MANU SIGH
Ventilation parameters
tidal volume 50~1500ml
frequency 4~60bpm
I:E 2:1~1:8
trigger sensitivity -1.0~-0.1Kpa
PEEP 0~2.0Kpa
pressure control 1~6Kpa
SIGH 1.5 breath times
breath pause 0~50%
Ventilation monitoring parameters
tidal volume. ventilation. IPPV ratio. breath /expiratory ratio. peak pressure . pressure – time wave. velocity-time wave. total respiratory rate. airway pressure wave. PEEP. breath pressure trigger. breath pause.
Monitoring oxygen concentration 21%~100%
Security Alert System
Oxygen Concentration alarm
upper limit setting range 50%~100%
low limit setting range 15%~50%
Airway pressure alarm
upper limit setting range 1~6Kpa
low limit setting range 0.4~2Kpa
Alarm minute ventilation
upper limit setting range 3.0~30L/min
low limit setting range 1~10L/min
Asphyxia alarm
the machine will give sound and light alarm if there is no
tidal volume input in 15s
Power failure alarm
Power AC 220V±10% 50Hz±1Hz
UPS. Battery. (choose and match)
Manipulators (choose and match)
Multi-function ICU Anesthesia Machine