Features
12.1″ high-resolution TFT color display USA BRAND
Simultaneous acquisition of 7 channel ECG waveforms displayed on one screen
Enlarged vital sign values to facilitate viewing from distance
ST-segment automatic analysis
100 groups of VPC data storage and arrhythmia waveforms and automatic analysis
Max. 7 hours ECG data and waveform storage after power-off
24 hours real time ECG waveform storage and replay
12/24/96-hour trend graph and data storage
Network capability
Optional IBP, CO2 (mainstream or sidestream) and built-in thermal recorder
Specifications:ECG | |
HR range | 0-300bpm |
Resolution | 1bpm |
Accuracy | ±1% or ±2bpm, Choose the higher one |
Bandwidth for ECG | 0.5Hz-75Hz Monitoring mode: 0.5Hz-40Hz; Diagnosis mode: 0.5Hz-75Hz |
Lead selection | standard I, II, III; aVR, aVL, aVF; V1-V6 |
ST segment | -2.0-2.0mV |
Arrhythmia analysis | Yes |
Lead-off alarm | Audible and visual |
Sweep speed | 12.5, 25, 50mm/sec |
Gain selection | X1/2, x1, x2, x4 and Auto |
Protection | Prevent AC4000V/50Hz voltage in isolation and work against electrosurgical interference, AC, EMG and defibrillation |
Alarm | High/Low limits, automatic saving |
NIBP | |
Method | Intelligent oscillation |
Parameters | SYST/DIAS/MEAN |
Operation modes | Manual/Auto/STAT (5 minutes) |
Pressure unit | mmHg/kPa |
Measurement range | Adult: 10-270mmHg Infant: 10-200mmHgNeonate: 10-135mmHg |
Range of cuff pressure | 0-300mmHg |
Cuff inflation | ≤10 secs |
Cuff deflation | ≤20 secs |
Alarm | High/Low limits, automatic saving |
RESP | |
Method | Chest impedance |
Measure Range | 0-150rpm |
Accuracy | ±2rpm |
Alarm | High/Low limits, automatic saving |
TEMP | |
Range | 0-50 |
Resolution | 0.1 |
Accuracy | ±0.2 |
Probe | Standard: Permanent dermal temperature probeOptional: Esophageal or rectal temperature probe |
Alarm | High/Low limits, automatic saving |
SpO2 | |
Display | SpO2 value, pulse histogram, waveform, pulse |
Range of SpO2 | 35%-100% adult/infant/neonate |
Resolution | 1% |
Accuracy | ±2%(75%-99%) ±3%(50%-75%) ±6%( 40%-50%) |
Range of pulse rate | 30-240bpm |
Accuracy of pulse rate | ±2bpm |
Sensor | Standard: finger tipOptional: infant Y-type clip, neonate wrap |
Alarm | High/Low limits, automatic saving |
CO2 | |
Measurement mode | Side-stream |
Parameters | Respiration rate, EtCO2, InsCO2 |
Resolution | ±2mmHg |
Alarm range | 0mmHg-120mmHg |
EtCO2 alarm default setting | 42/20mmHg |
InsCO2 alarm default setting | 10/0mmHg |
Sampling rate | 60 times/second |
Temperature for use | 10-40 |
Temperature for storage | 0-65 |
Relative moisture for use and storage | 5%-95% |
Atmospheric pressure range | 500mmHg-800mmHg |
IBP | |
Measurement mode | IBP transducer |
Transducer sensitivity | 5μV/V/mmHg |
Measurement range | -30mmHg-300mmHg |
Parameters | SYST, DIAS, MEAN |
Calibration mode | Zero and 100mmHg calibration (optional) |
Resolution | 0-100mmHg ±4mmHg |
100mmHg-300mmHg ±2% | |
Measurement sites | ART, PA, CVP, RAP, LAP, ICP |
Sampling rate | 512 times/second |
Temperature for storage | 0-65 |
Recorder | |
Recorder type | Built-in thermal recorder |
Recording mode | Recording waveform, text in real time |
Configuration | |
Display | 12.1″color TFT display |
Standard | Three-channel ECG, NIBP, SpO2, RESP, TEMP, Pulse Rate, Built-in battery |
Optional | IBP, CO2 (mainstream or sidestream), built-in thermal recorder |
Others | |
Power supply | AC100-250V, 50/60Hz |
Power consumption | ≤80VA |
Battery | Rechargeable sealed lead acid battery |
Operating time | 2 hours or more |
Trend time | 1-96 hours |
Net/Gross weight | 6.5kg/11.5kg |
Packaging | 500L X 320WX 460H (mm) |
Safety standard | IEC601-1 Class II, Type BF, CF |
Multi-parameter patient monitor