Working Principle
The SHOWLAND-2012 laser particle size analyzer is based on full measuring range of Mie scattering theory, sufficiently considering the optics nature such as sample and decentralized medium refraction coefficient and so on. On the basis of laser illuminate size and the size distribution regular pattern of scattering light energy that comes from the sample inversely calculates out sample group.
Technique Parameter
1. Measurement range:0.1-600 u m
2. Error:< ± 1% ( national standard materials D50)
3. Repetitiveness deviation:< ± 1% ( national standard materials D50)
4. Electrical requirement:AC 220 ± 10 V, 50 Hz, 200w
5. Operating System: Windows2000, XP
6. Connecting Port: USB
7. Dimensions:950 X 330 X 300 mm
8. Net Weight:40 kg
Software Introduction
PADMAS is the laser particle size analyzer measurement analytic system (Particle Diameter Measure & Analysis System)abbreviation. The software is continuously improved and upgraded. It has become the excellent in the present size analysis instrument profession.
Apparatus composition
Apparatus light route and the probe is patent designed, scattering light can be collected at best. The largest test scattering angle is up to 135°. The quartz claimed sample window is more wear than the ordinary sample window. The inside cavity is completely spouted blackly; reflected light is utmostly sucked up, decreasing diffuse reflection. Light route and probe are specially manufactured. Probe accuracy and dependability are gurranted.
FOR MORE INFO./OTHER DETAILED DEMANDS, PLEASE CONTACT WITH US FREELY.
Laser Particle Size Analyzer