1MAR-100Multi-Purpose Infusion Controller 2output fluid temperature is up to 2-5degree3220V~ 50Hz/60 Hz
Detail information
Technical specification and Functions
Product | Multi-Purpose Infusion Controller ---Infusion Guard |
Model | MAR-100 |
Tube | 15 drops/ml, 20 drops/ml, 60 drops/ml |
Principle | Gravity Infusion; Infrared photoelectric beam detects drips; controlling valve automatic controls infusion rate; and heating function. |
Operational condition | Ambient temperature : +5°C~40°C Relative humidity : 20%~80% |
Power source | Both AC. and DC.power supply 220V~ 50Hz/60 Hz 7.2V 800mAH |
Power consumption | 3 VA (without heating) 12VA (heating) |
Range of setting | Drop rate: 1drop/min ~ 140 drops/min Flow rate : 15 drop/ml: 4 ml/hour ~ 560 ml/hour; 20 drop/ml: 3 ml/hour ~ 420 ml/hour; 60 drop/ml: 1 ml/hour ~ 140 ml/hour. Volume limit: 1ml ~ 999ml (over 999 ml displays as“all”) Time setting : 1min ~ 23 hours and 59mins Total volume: 1ml ~ 9999ml |
Accuracy | Drip rate : ±3% ; Flow rate :±5% ; Time : ±5% |
Heating | output fluid temperature is up to 2°C~5°C increase in the low-temperature environment. |
Alarm | Sound and light alarm when battery disconnection, low battery, Finish, drops-checking failed and leakage |
Dimensions | 135mm × 110mm × 45mm |
Weight | 400g(battery includes) |
MAR-100Multi-Purpose Infusion Controller