UV/VIS Spectrophotometer
Model: UV-9200
FEATURES:
1 Microprocessor control, 16×2 LCD display;
2 Auto zero and auto 100%T adjustment provided;
3 Calibration curve can be set up by either measuring or entering up to 10 standards or entering K and B factors directly via the
keyboard;
4 Parameters can be saved for later use; Up to 10 calibration curves can be stored and edited for user's convenience;
5 Standard manual 4-cell holder accommodates cells from 5-50mm and changeable to long path length cell holder of 100mm.
6 Data can be printed out via a printer port;
7 PC control measurement can be achieved via a USB port for more accurate and flexible requiements.
SPCIFICATIONS:
Model | UV-9200 |
Wavelength | 190~1100nm |
Wavelength accuracy | ≤±2.0nm |
Wavelength reproducibility | ≤1nm |
Monochromator | Single-beam, C-T type, grating 1200L/mm |
Photometric accuracy | ≤±0.5%T |
Photometric reproducibility | ≤0.3%T |
Stray light | ≤0.1%T (Nal at 220nm, NaNO2 at 340nm) |
Spectral bandwidth | 2nm (1, 4, 5nm optional) |
0%T stability | ≤0.2%T(30min) |
100%T stability | 0.001A/30min (at 500nm, after warming up) |
Operation mode | T, A, C, E |
Photometric range | -0.3~3A |
Display | 16X2 LCD |
Detector | Silicon Photodiode |
Light source | Tungsten halogen lamp, D2 lamp |
Power Requirement | 110~240V,50/60HZ |
Power consumption | 120W |
Dimensions | 530×410×210mm |
Net weight | 16kg |
UV-9200 UV/VIS Spectrophotometer