Welcome to ECVERY
Cleveland Open Cup Flash Point Tester (Digital Display)
Cleveland Open Cup Flash Point Tester (Digital Display)
Supplier Info
[China Supplier]
Contact Person : Mr. Shi Steven
Tel : 86-21-24287122
Fax : 86-21-54792677
Product Detail
Cleveland Open Cup Flash Point Tester -ASTM D92 and ISO2592 -It is newly design,having small structure

 Cleveland Open Cup Flash Point Tester

Model: SYD-3536-1

Application

This instrument is designed and made as per standards GB/T3536-2008 Test Method for Flash Point and Fire Point of Petroleum Products - Cleveland Open Cup Method,ASTM D92 and ISO2592. It is suitable to determine the flash point and fire point of petroleum products which flash points higher than 79 °C except fuel oils.

I. Main technical features

1.It adopts technology of single chip machine and LCD screen. English man-machine interface.The heating device is continuously stepless adjustable.Easy to set parameters.

2.The LCD screen has prompt menu, prompt type input for operation interface. It shows set parameters and real-time display sample temperature and other parameters. Press the record key when flash point appearing. The screen will display and save flash point value.

3.It is newly design,having small structure. Stainless steel working table make it elegant and beautiful. The operator can do determination as long as connecting with coal gas or other civil gas. Easy to use and results are exact.

4.It is equipped with wind-shelter and flame extinguishing cover which are accord with requirements of test.

5.Accurate heating rate.The instrument can do test automatically. Operator only need to observe the flash point appearing. The cost performance is high.

II. Main technical specifications and parameters

1.Power supply: AC(220±10%)V,50Hz.

2.Heating device: Electric furnace heating, no naked fire, explosion prevented.The power is adjustable from 0W∼600W. The max heating temperature can reach 400°C.

3.Temperature control: Single chip machine.The heating rate can meet requirement of standard GB/T 3536-2008

4.Temperature display: LCD screen shows temperature parameters. The range is 0°C,~400°C, accuracy is 0.1°C.

5.Flame application device: scanning the flash point automatically.

6.Temperature sensor: PT100 RTD

7.Ignition device: (1) Ignition source: coal gas(or civil gas,the same below)

(2) Bore diameter of spout: 0.8mm

8.Ambient temperature: (-10∼50)°C

9.Relative humidity: ≤85%

10.Overall power consumption: ≤ 650W

11.Overall dimension: 340mm*320mm*450mm (L*W*H, Temperature sensor is included)

 

 

 

Cleveland Open Cup Flash Point Tester (Digital Display) ,Digital Cleveland Open Cup Flash Point Testing Machine,Cleveland Open Cup Flash Point Appatatus(Digital Display)

Cleveland Open Cup Flash Point Tester (Digital Display)

Ads by Google

About Us | Contact Us | Help | Terms & Conditions
Hot Products: A | B | C | D | E | F | G | H | I | J | K | L | M | N | O | P | Q | R | S | T | U | V | W | X | Y | Z | 0-9
Copyright Notice @ 2008-2022 ECVERY Limited and/or its subsidiaries and licensors. All rights reserved.