The flame photometer is an analytical instrument based on the principle that the emission spectrum decides the element.

Flame Photometer FP640

Major Technical Specifications

     a. Electricity supplier:  photocell;

     b. Prismatic Measure:  interfering filter;

     c. Indication :  double-channel digital indication;

     d. Sensitivity  :   in clinical measurement;

                             K: 0.0-9.9mmol/L;

                             Na: 0-199mmol/L;

     e. Repeatability  : Cv≤2%

     f. Linear Disparity  :   ≤5% when K varies from 0.02 to 0.06mmol/L and Na varies from 1.00 to 1.80mmol/L.

1.2.2 Criteria for Normal Usage

     a. Temperature       : 10~35°C;

     b. Prismatic Measure  : ≤80%

     c. The instrument should be laid flat, and kept away from the sunlight, EMI or oscillation;

     d. Fire extinguisher should be equipped, and good ventilation is needed;

     e. Electric Power Voltage: 220±22V; frequency: 50±0.5Hz; It should be well grounded.

1.2.3 Instrument volume, weight and package.

     a. Volume  : 400*250*500mm;

     b. Weight  : 20kg

     c. Set     : main body, air compressor, gasoline gasifier, liquidated petroleum container and accessories.