Ads by Google
Technical data | |
Power consumption | ≤30VA |
Dimension | 230L × 110W × 240H (mm) |
Net weight | ≤3.5kg(without packing) |
Electrical standard | Class I |
Electrical type | Type BF |
Power requirement | Main power; AC 160~AC 265 V, 5060/ Hz |
internal power | NiMH battery 12V(fully charged, ≥4 hrs at 25ml/h) |
Operating environment | |
Temp. | +15°c~40°c |
Humidity | 20%~90% |
Atmosphere pressure | 70~106 kPa |
Altitude | ≤2000m |
Transportation and storage | |
Temp. | -20°c~55°c |
Humidity | ≤93% |
Atmosphere pressure | 50~106 KPa |
No corrosive gas, ventilative | |
Performance parameter | |
Recommended infusion sets | Standard silica gel infusion sets, 20 drop/ml |
Flow rate setting scale | 1.0~1200 ml/h(in 0.1 ml/h increment) |
Bolus rate | 600/800/1000/1200ml/h for option |
Prime rate | 600/800/1000/1200ml/h for option |
KVO rate setting scale | Off.1.0~5.0 ml/h(in 0.1 ml/h increment, auto control) |
Max volume limit | 9999 ml(min increment 0.1 ml/h) |
Max volume limit accumulated | 9999 ml(min increment 0.1 ml/h) |
Accuracy | ±5% |
Occluded alarm threshold | 5 level(300/500/700/900/1100 mmHg) |
Bubble detective sensitivity | Single bubble volume ≥ 0.03 ml is detectable |
Alarm | Failure, Low battery, AC failure, Door open, Occlusion Air in line, Infusion finished(auto switch to KVO) |
IV infusion pump (manufacturer with CE)