100ma bedside x-ray machine with CE Max 100kv
I. General Description:
Type SF100BY mobile diagnostic X-ray machine with section combined X-ray tube head, full wave rectification, is suitable for all kinds of hospitals making x ray and photographing, especially for injure dept. of orthopedics, operating room and household ward.
II. Advantages:
This efficiency-improved x ray machine takes the following advantages:
1. Adopts high-pressure silicon, full-wave rectification technology.
2. With kilovolt (kV), milliampere (mA) and time (s ) three factors interlocks control system
3. Overloading indicator & the protection circuit of excessive voltage & insufficient voltage, thus the
X-ray tube head of x ray will not be damaged because of mis-operating.
4. Equipped with steady filament voltage circuit to control the X ray dosage while photographing.
5. Remotely controlled exposure & manually controlled exposure options
6. Sound & flash indicator while exposing for patients & medical personnel convenience
7. This machine equipped the dynamoelectric cluster limiter with shadow positioning.The electric cluster limiter will send out a beam of visual light while operating to aim the nidus and the scale of the visual light overlaps the scale of X ray exposing. This function
make it convenient that medical personnel position the nidus & the x ray exposed area.
8. This machine has a movable cantilever sliding in vertical direction so that it is flexible to adjust the Xray tube head to appropriate angle & height. When the complete machine is moved, the cantilever
of the Xray tube head can be set to the minimum position and lock, safely and conveniently.
9. The gross weight of the complete machine is150 kilograms, width is 72cm, high is
200cmand length is 118cm. It can be used to moved in the ward and between beds conveniently.
III. Main Technical Conditions
Maximum Ratings:
S
kV mA | ~3.2 | ~4.0 | ~6.3 |
16 | 90 | 90 | 90 |
32 | 90 | 90 | 90 |
63 | 90 | 90 | |
100 | 80 |
3. Condition of the power supply
Single-phase alternating current
Voltage: 180V-250V ; Frequency: 50Hz ; Power supply resistance ≤0.2Ω
Instantaneous current: 37A (Max)
4. Timer 0.08-6.3s, 20 shelves of electron type
5. X-ray tube specification : XD4-29/100, focus 4.3
6.X-ray tube focus highest position (above ground level)≥1710mm X ray
tube focus lowest position (above ground level)≤480mm
7.Remotely exposure distance: 6m
8. The required pulling force to move the whole machine≤250N (25kg)
9. Power of Electric Beam Limiting Device: AC 24V, Capacity 110W.
10.Indicator Power of Electric Beam Limiting Device: AC 24V, Power: 100W, Tungsten Haloid Light
11. Illuminance of Electric Beam Limiting Device: no less than 100Lx at 1000mm from
X-ray focus to target area
12. When Xray focus at 650mm away from the film, max illumination scale ≤350mm×
350mm; When Xray focus at 1000mm away from the film, max illumination scale
≤50mm×50mm
100ma bedside x-ray machine with CE Max 100kv