Product Detail
IEC60695-11-3/4 Horizontal and Vertical Flame Tester 1.Reference to IEC60695-11-3/4,UL94 2.Touch screen control

IEC60695-11-3/4 Horizontal and Vertical Flame Tester

1.Reference to IEC60695-11-3/4,UL942.Touch screen control

3.One button operation

1.Introducation:

 When there is a fire around the products, should ensure that it will not to cause the spread of the fire, horizontal-vertical burning test simulates electronic products around the environment in the early condition, using the simulation technology to assess the fire risk. The test equipment apply to the electrical and electronic product components of household appliances, spare parts and components, such as: household appliances insulation shell, Panel switches, printed circuit boards and insulation materials and so on.2.Characteristics:

Horizontal and vertical flame chamber tester is composed by the test part and control part, using the integrated design, to facilitate the installation at the scene and debug. The shell of the test chamber and the important parts are all manufactured by stainless steel, resistant cigarettes and gas corrosion. The automatic degree of control system is high. Has automatic ignition, the products which were tested in both vertical and horizontal orientation has the function automatically follow flame in the cross of the test. Time and temperature figures show, so that the observation and records is facilitate and the use of it is stable and reliable.

 3.Technical Paremeters:

Kinds

  500W(5V class ) 50W(HB/Vclass)
Test Flame500W(burning test for V-0,V-1,V-2  classification)

50W(burning test for V-0,V-1,V-2 classification)

Burner TubeInternal diameter: Φ 9.5mm±0.3mm

Internal diameter: Φ 9.5mm±0.3mm, 

Burning Angle0°,20°,45°

0°,20°,45°

ThermocoupleType K (Ni/Cr - Ni/Al)

Type K (Ni/Cr - Ni/Al)

Thermocouple Size 0.5 mm

0.5 mm

Standard Copper Block Φ9mm±0.01mm,weight:10g±0.05g before drilling.  

Φ5.5mm±0.01mm, weight.:.76g±0.01g before drilling. 

Time for Temperature to IncreaseFrom 100°C±5°C to 700°C±3°C; 54s±2s (preset)

From 100°C±5°C to 700°C±3°C;44s±2s (preset)

Gas Flow965±32ml/min

105±10ml/min

Gas Supplymethane,min.98% purity

methane,min.98% purity

water-column 125m±5mm

‹10mm

Test  Time54±2s

44±2

Flame Height Indicator40±1mm

20±1mm

Volume Of Chamber>0.5m³, black interior

>0.5m³, black interior

Outer DimensionsW*D*H=1200mm*700mm*1400mm

W*D*H=1200mm*700mm*1400mm

Conform to standardIEC6065-11-3,UL94, IEC60695-11-10

UL94,IEC60695-11-4,IEC60695-11-10

 

( function(){ var aBigImageSrc = [ 'http://i00.i.aliimg.com/photo/v1/702832130_1/IEC60695_11_3_4_Horizontal_and_Vertical.jpg' , 'http://i00.i.aliimg.com/photo/v1/702832130_2/IEC60695_11_3_4_Horizontal_and_Vertical.jpg' , 'http://i00.i.aliimg.com/photo/v1/702832130_3/IEC60695_11_3_4_Horizontal_and_Vertical.jpg' , 'http://i00.i.aliimg.com/photo/v1/702832130_4/IEC60695_11_3_4_Horizontal_and_Vertical.jpg' ]; var oProductShow = new AE.run.minisite.productShow(); oProductShow.init( { 'aImgSrc': aBigImageSrc } ); } )(); IEC60695-11-3/4 Horizontal and Vertical Flame Tester