The Specification of MA-900 | |||||
Product name: | Multi-functional infusion pump | ||||
Model: | MA-900 | ||||
Pumping type: | Peristaltic finger pump | ||||
IV set requirement: | Standard IV set (15,20 drops/ml), or micro drip IV set (60 drops/ml) with 3.4~4.5mm tubing | ||||
Flow rate setting range: | 1 ml/h~999 ml/h | ||||
Drip rate setting range: | 1 drop/min~333 drops/min | ||||
Flow rate step: | 0.1 ml/h | ||||
Flow rate accuracy: | ±5% ( standard PVC set must first be adjusted) | ||||
Drip rate accuracy: | ±1.5 % | ||||
Timing range: | ±30% | ||||
Volume setting range: | 0~9999ml | ||||
Volume infusion display range: | 0~9999ml | ||||
Purge rate: | 800 ml/h | ||||
KVO (keep vein open): | 1 ml/h | ||||
Air bubble detection method: | Ultrasonic | ||||
Drip inspection method: | Infrared photo-electric beam | ||||
Peristaltic finger pump equipment: | Detachable | ||||
Display: | Color LCD with high backlight | ||||
Block pressure detector: | 40KPa~140KPa, sensitive adjustment of 8 levels | ||||
Alarms: | Infusion Monitoring: | ||||
Air in tube, door open, occlusion, pause, infusion finishing, abnormal dripping,leakage, tube slope etc. | |||||
Diagnosis malfunctions: | |||||
Low battery,damaging battery,abnormal dropping sense organ,abnormal electrical set. | |||||
AC power source: | 100V~240V, 50Hz~60Hz | ||||
Inner battery: | Rechargeable Ni-hydrogen battery, DC12V, 2000mAh | ||||
Charge time: | 16 hours | ||||
Discharge time: | ≥5 hours | ||||
Power consumption: | AC power supply: <20 VA, battery supply: <5 W | ||||
Weight: | 2.1 kg | ||||
Safe classification: | Class 2;type CF; Portable; IPX 4 | ||||
Dimensions: | 140mm×168mm×230mm | ||||
Operational condition: | Temperature: +5°C~+40°C; Relative humidity: ≤80% | ||||
User range: | Venous infusion,Vasodilator medicianl preparation,Anticancer drug,Accelerative procreation drug, Nutrient liquid and other kinds of medicinal liquid , mini infusion and rapid infusion. |
Multi-functional infusion pump